کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
2507694 | 1117501 | 2015 | 9 صفحه PDF | دانلود رایگان |
A selective and rapid high-performance liquid chromatography–tandem mass spectrometry method was developed and validated for the quantification of raltegravir using raltegravir-d3 as an internal standard (IS). The analyte and IS were extracted with methylene chloride and n-hexane solvent mixture from 100 µL human plasma. The chromatographic separation was achieved on a Chromolith RP-18e endcapped C18 (100 mm×4.6 mm) column in a run time of 2.0 min. Quantitation was performed in the negative ionization mode using the transitions of m/z 443.1→316.1 for raltegravir and m/z 446.1→319.0 for IS. The linearity of the method was established in the concentration range of 2.0–6000 ng/mL. The mean extraction recovery for raltegravir and IS was 92.6% and 91.8%, respectively, and the IS-normalized matrix factors for raltegravir ranged from 0.992 to 0.999. The application of this method was demonstrated by a bioequivalence study on 18 healthy subjects.
Journal: Journal of Pharmaceutical Analysis - Volume 5, Issue 2, April 2015, Pages 101–109