کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2964117 1178638 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Defibrillation efficacy of a subcutaneous array lead: A case report
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Defibrillation efficacy of a subcutaneous array lead: A case report
چکیده انگلیسی

SummaryWe report a case of Brugada syndrome with a high defibrillation threshold (DFT) in whom a subcutaneous array lead was used to lower the DFT in combination with a transvenous right ventricular defibrillation lead. The patient had previously received pacemaker implantation due to sick sinus syndrome. An implantable cardioverter defibrillator (ICD) with a transvenous right ventricular defibrillation lead alone required a high DFT. A subcutaneous array lead improved defibrillation efficacy in combination with a right ventricular lead. These data suggest that a subcutaneous array lead facilitates implantation of an effective ICD lead system in patients requiring a high DFT.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cardiology Cases - Volume 1, Issue 1, February 2010, Pages e21–e24
نویسندگان
, , , , , , , , , ,