کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4226997 1609812 2010 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of left atrial function by multidetector computed tomography before left atrial radiofrequency-catheter ablation: Comparison of a manual and automated 3D volume segmentation method
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Evaluation of left atrial function by multidetector computed tomography before left atrial radiofrequency-catheter ablation: Comparison of a manual and automated 3D volume segmentation method
چکیده انگلیسی

IntroductionThe purpose of this study was to compare a manual and automated 3D volume segmentation tool for evaluation of left atrial (LA) function by 64-slice multidetector-CT (MDCT).Methods and materialsIn 33 patients with paroxysmal atrial fibrillation a MDCT scan was performed before radiofrequency-catheter ablation. Atrial function (minimal volume (LAmin), maximal volume (LAmax), stroke volume (SV), ejection fraction (EF)) was evaluated by two readers using a manual and an automatic tool and measurement time was evaluated.ResultsAutomated LA volume segmentation failed in one patient due to low LA enhancement (103HU). Mean LAmax, LAmin, SV and EF were 127.7 ml, 93 ml, 34.7 ml, 27.1% by the automated, and 122.7 ml, 89.9 ml, 32.8 ml, 26.3% by the manual method with no significant difference (p > 0.05) and high Pearsońs correlation coefficients (r = 0.94, r = 0.94, r = 0.82 and r = 0.85, p < 0.0001), respectively. The automated method was significantly faster (p < 0.001). Interobserver variability was low for both methods with Pearson's correlation coefficients between 0.98 and 0.99 (p < 0.0001).ConclusionsEvaluation of LA volume and function with 64-slice MDCT is feasible with a very low interobserver variability. The automatic method is as accurate as the manual method but significantly less time consuming permitting a routine use in clinical practice before RF-catheter ablation.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Radiology - Volume 75, Issue 2, August 2010, Pages e141–e146
نویسندگان
, , , , , , , , , , ,