Internally cooled antenna for microwave ablation: Results in ex vivo and in vivo porcine livers
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Internally cooled antenna for microwave ablation: Results in ex vivo and in vivo porcine livers
چکیده انگلیسی

PurposeTo evaluate the performance of microwave ablation with an internally cooled antenna in ex vivo and in vivo porcine livers.Materials and methodsMicrowave ablations were performed in ex vivo and in vivo porcine livers with a 1.9 mm-diameter 2450 MHz microwave needle antenna. Four power settings (50, 60, 70 and 80 W) were used in this study, while application time was fixed at 10 min. Diameters of the coagulation zone were observed on gross specimens.ResultsExcessive elongation of coagulation along the needle shaft was not encountered in all ablations. An arrow-shaped charring was observed. As the power increases, the long-axis coagulation diameter for in vivo liver increased significantly (P < .05). The short-axis coagulation diameter for in vivo liver was significantly smaller than that for ex vivo liver (P < .05) but not statistically different among the four power settings. After 10 min of ablation at 80 W, the short-axis and long-axis coagulation diameter for in vivo liver was 4.92 ± 0.15 cm and 2.37 ± 0.10 cm, respectively.ConclusionThe internally cooled microwave antenna may be advantageous to minimize collateral damages.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Radiology - Volume 67, Issue 2, August 2008, Pages 357–361