کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4229706 1610062 2015 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Feasibility of low contrast media volume in CT angiography of the aorta
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Feasibility of low contrast media volume in CT angiography of the aorta
چکیده انگلیسی

ObjectivesUsing smaller volumes of contrast media (CM) in CT angiography (CTA) is desirable in terms of cost reduction and prevention of contrast-induced nephropathy (CIN). The purpose was to evaluate the feasibility of low CM volume in CTA of the aorta.Methods77 patients referred for CTA of the aorta were scanned using a standard MDCT protocol at 100 kV. A bolus of 50 ml CM (Iopromide 300 mg Iodine/ml) at a flow rate of 6 ml/s was applied (Iodine delivery rate IDR = 1.8 g/s; Iodine load 15 g) followed by a saline bolus of 40 ml at the same flow rate. Scan delay was determined by the test bolus method. Subjective image quality was assessed and contrast enhancement was measured at 10 anatomical levels of the aorta.ResultsDiagnostic quality images were obtained for all patients, reaching a mean overall contrast enhancement of 324 ± 28 HU. Mean attenuation was 350 ± 60 HU at the thoracic aorta and 315 ± 83 HU at the abdominal aorta.ConclusionsA straightforward low volume CM protocol proved to be technically feasible and led to CTA examinations reaching diagnostic image quality of the aorta at 100 kV. Based on these findings, the use of a relatively small CM bolus can be incorporated into routine clinical imaging.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Radiology Open - Volume 2, 2015, Pages 58–65
نویسندگان
, , , , , , , , ,