An 80-year-old man, who had dilated cardiomyopathy with right ventricular (RV) dilatation, underwent implantable cardioverter defibrillator (ICD) implantation for advanced atrioventricular block and primary prevention of sudden cardiac death. Tined and screw-in leads were placed on the right atrial appendage and RV apex, respectively. Ventricular pacing inhibition was detected after surgery due to oversensing by diaphragmatic myopotential occurring only during deep inspiration. We performed re-surgery and switched the screw-in lead for a tined lead. The diaphragmatic myopotential decreased, thereby improving oversensing by diaphragmatic myopotential and ventricular pacing inhibition. It might be beneficial to use a tined lead when placing the ventricular lead at the RV apex for implantation of a pacemaker or ICD if oversensing of diaphragmatic myopotential is observed using a screw-in lead.