Quantifying the performance of two different types of commercial software programs for 3D patient dose reconstruction for prostate cancer patients: Machine log files vs. machine log files with EPID images
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Quantifying the performance of two different types of commercial software programs for 3D patient dose reconstruction for prostate cancer patients: Machine log files vs. machine log files with EPID images
چکیده انگلیسی
For patients, differences in dosimetric parameters were within 1% for almost all fractions. PerFRACTION had wider range of dose difference between first fraction and the other fractions than Mobius3D (e.g., maximum difference: 0.50% for Mobius3D vs. 1.85% for PerFRACTION), possibly because EPID may detect some types of MLC positioning errors such as miscalibration errors or mechanical backlash which cannot be detected by log files, or that EPID data might include image acquisition failure and image noise.