Possibilities and limitations of quadrupole mass spectrometric detector in fast gas chromatography
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Possibilities and limitations of quadrupole mass spectrometric detector in fast gas chromatography
چکیده انگلیسی
In this work the application and limitations of a common bench top quadrupole mass spectrometer was evaluated for the qualitative and quantitative measurement of n-alkanes and pesticides of a wide range of volatilities and polarities with fast GC separations using 0.15 mm I.D. narrow-bore capillary columns. It was found that the spectra acquisition rate has a great impact on sensitivity (peak areas, peak shapes and S/N ratios). The quality of the obtained spectra is not significantly influenced in the full scan monitoring mode for the fastest scan rates. For quantitative analysis a selected ion monitoring mode is able to acquire the sufficient number of data-points for the proper peak shape reconstruction and good repeatability of peak areas measurements expressed by RSD (<5%) for all tested dwell times shorter than 75 ms. However, for shorter dwell times, S/N ratio is lower, while peak areas are not influenced.