Evaluation of particle size distribution of albuterol sulfate hydrofluoroalkane administered via metered-dose inhaler with and without valved holding chambers
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی امراض پوستی
پیش نمایش صفحه اول مقاله
Evaluation of particle size distribution of albuterol sulfate hydrofluoroalkane administered via metered-dose inhaler with and without valved holding chambers
چکیده انگلیسی
In vitro results obtained under these test conditions demonstrate that all the FPM values for the VHCs tested were within 15% of the control range, a difference that is unlikely to be clinically meaningful. These results do not warrant a change in the recommended dose of albuterol hydrofluoroalkane administered when using the VHCs tested. The use of an MDI in conjunction with a VHC provides a reasonable therapeutic approach for administration of albuterol hydrofluoroalkane to young children and other patients who have difficulty administering the MDI alone.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Annals of Allergy, Asthma & Immunology - Volume 94, Issue 1, January 2005, Pages 80-85