کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9748973 1493797 2005 12 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Application of capillary zone electrophoresis and reversed-phase high-performance liquid chromatography in the biopharmaceutical industry for the quantitative analysis of the monosaccharides released from a highly glycosylated therapeutic protein
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Application of capillary zone electrophoresis and reversed-phase high-performance liquid chromatography in the biopharmaceutical industry for the quantitative analysis of the monosaccharides released from a highly glycosylated therapeutic protein
چکیده انگلیسی
Two assays for the quantitative determination of the neutral and amino-monosaccharides attached to a therapeutic glycoprotein were developed using capillary zone electrophoresis (CZE) and RP-HPLC. These assays meet the strict batch release requirements of the quality control in biopharmaceutical industry. The monosaccharides were released from the glycoprotein by hydrolysis with 2N trifluoroacetic acid. In the CZE assay the monosaccharides were reacetylated prior to derivatization with 8-aminopyrenesulfonic acid (APTS), reacetylation in the glycoprotein matrix was investigated in detail. The RP-HPLC method used pre-column derivatization with anthranilic acid in methanol-acetate-borate reaction medium; reacetylation was not necessary. However, epimerization of the different monosaccharides was observed and studied in detail. For the quantitative assay, separation of the amino-monosaccharide epimers had to be developed. The HPLC assay was validated.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1079, Issues 1–2, 24 June 2005, Pages 354-365
نویسندگان
, , ,