کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4725239 1355971 2009 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
The current performance of the in situ14C extraction line at ETH
موضوعات مرتبط
مهندسی و علوم پایه علوم زمین و سیارات ژئوشیمی و پترولوژی
پیش نمایش صفحه اول مقاله
The current performance of the in situ14C extraction line at ETH
چکیده انگلیسی

We present the new 14C extraction line at ETH Zürich. This system is designed to extract in situ-produced cosmogenic 14C from terrestrial quartz samples, and to obtain pure CO2 gas for analysis with a gas ion source Accelerator Mass Spectrometry (AMS) system. Samples are degassed at 1550–1600 °C without the use of a fluxing agent. Gas purification is achieved by a series of cryogenic traps and passage through hot Ag and Cu wool/mesh. Graphitization and, thus, sample dilution is not required. Tests to determine the CO2 recovery after gas extraction and cleaning yielded consistently good recovery rates of >99.8% (n = 7). The 14C blank contribution from the all-metal tubing system is negligible. Our preliminary procedural blank estimate – deriving mostly from the hot extraction furnace – is <5 × 10514C atoms. Extraction tests on two quartz samples by stepped-heating show a quantitative separation of atmospheric 14C at ≤500 °C from the in situ component above 1200 °C. Based on these data, we estimate to achieve a complete 14C extraction from a quartz sample.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Quaternary Geochronology - Volume 4, Issue 6, December 2009, Pages 493–500
نویسندگان
, , , , , ,